
3000.doc Iss. 2/00
4
TECHNICAL SPECIFICATION
DISPLAY: LCD Graphics Panel, 240 x 64 dot, with LED backlight contrast control via
front panel keyboard.
INPUTS: 2 channels for PRTs via 6 pin Lemo sockets. LEMO type number for input
plug: FGG1B.306.CLAD52
IMPUT IMPEDENCE > 10MΩ
Max. input Voltage ± 40Vdc, 28Vrms
Common Mode Rejection 50Vrms
The inputs are linearised and user selectable to the following standards:-
IEC751 R0=100Ohm, Ra=3850
US/JIS R0=100 Ohm,Ra=3851
EN-751 R0=100 Ohm, Ra=3850 in accordance with ITS90 standards
USER Up to 20 probes may be configured to customer entered constant, A
B, C, and R0 values.
A Standby current is always passed through the sensors.
UNCERTAINTY OF MEASUREMENT:
INPUT RANGE
°C
RESISTANCE
Ohm
CURRENT RESOLUTION
°C °F K
UNCERTAINTY
1 year @ 20 ±5°C
Pt100 -200 … -100 18 … 60 1mA 0.001 0.02°C
-100 … +500 60 … 280 1mA 0.001
0.01°C
+500 … +800 280 … 450 1mA 0.001
0.02°C
2 channels for thermocouples (Model 3001) via 4mm sockets. A range of
thermocouple plugs with binding posts is available as accessories , see price
list.
The Inputs are linearised for the following thermocouple Types
B,E,J,K,N,R,S,T and in accordance with the following Standard, NIST 175,
ITS 90. Types C & D to ASTM E988
Reference Junction compensation may be selected for the following modes:-
Automatic:- Internal Reference Junction range 0 to +40°C
External:- Via Pt100 sensor connected to channel A or B range 0 to
+100°C
OFF:- Turns the Reference Junction OFF = 0°C
Comentarios a estos manuales