3000.doc Iss. 2/00
12
The display will first show the count down timer once the start time is reach the display will indicate
run and the number of scans will be display. Once the set number of scans has completed the display
will return to the Hold position until triggered again.
DATA LOGGING SETUP
To store values enter the MEM option and select Log select the Set Up option and the number of
samples or readings to be stored will be displayed. To accept this number press Yes, to change the
number press No, and then enter the number of samples you wish to store, (this number must be
between 1 and 4000.) Press OK, you will then be asked to confirm your choice. Press Yes, the
logging mode can switched on or off and the previous stored values scrolled to the display using the
Rev option. The values will be displayed with log number, channel measured, the temperature value,
date and time.
Fig.9 : Data logging Menu Fig.10: Screen showing logged values
To scroll the pages use the left hand up/down arrow keys, and to exit the review screen press the
OK/MEM key.
To clear the current logged values select New and the display will ask you to confirm that you wish to
clear the current stored values. Press Yes or No. If Yes is pressed you will be requested to confirm
your choice, press Yes to confirm. The logging mode may then be activated by pressing the On key.
When the set-up is complete press Quit and return to the measurement mode. The bottom line of the
display will show the data log status information:-
“Hold Samples = 0/20 ”
To trigger the measurements and store samples press the TRIG key. This line will now change to
“Run Sample=#/20”, where # = the number of samples stored. When the set number of values are
stored, the measurements return to the Hold state. Press the TRIG key again and the display will ask if
you wish to clear the current log. Press Yes and confirm, the TRIG key will now start a new measure
and store cycle.
The values logged will always relate to the channel on the display. It is possible to stop the logging
function, change channels and restart logging using the TRIG key (see TRIG section below).
A0 Pt100 EN60741 R0 = 100.000 a = 3851 4wre
+ 100.000
°
C
Input =138.500 Ohm +I Busy
Run SL1 0/10